Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Rahul Kashyap
0

 

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)

Procedure:

Cleaning:

·         Check that the power supply to the instrument Should be switched OFF before cleaning.

·         Clean the Instrument with a clean dry cloth every day.

·         Clean the UATR properly after each analysis.

Calibration of Fourier Transform Infrared Spectrophotometer (FTIR)


Operating Instructions:

·         Put on the main power supply of instrument.

·         Switch on the instrument and computer attached to the instrument taking care that while switching on the instrument only normal base accessory shall be placed in the instrument.

·         Double click on the spectrum icon to open the software.

·         Feed your user name and password and press enter.

·         Click on < Connect> to Connect the Instrument.

Background Selection:

·         Go to <Measurement Menu> Select< Background> to start background.

·         After completion of background, prepared standard and sample as per standard test procedure.

Preparation of Sample table (Using UATR accessory):

·         Go to<View> Select <Data explorer>.

·         From the left side assistant bar select sample table and fill the parameters regarding to sample ID and description.

·         Select the data Path location from the top of sample table by click on browse.

·         Go to left side assistant bar and select sample view.

·         Go to <set up menu> Select < instrument>. A set up window is open in bottom side.

·         Go to < Set up instrument basic> and set the parameter as per requirement. Then <load & Save>. Click <Save current setting> and press ok.

Scanning for Solid samples

·         Place the Respective Standard & Sample on the crystal in the sequence of sample

Table.

·         Give the pressure 30 to 40.

·         Go to sample table and select standard as well as sample which put on the crystal.

·         Go to <measurement> Select <Scan> to Started Scanning.

·         Click on sample table on top right side and click to save.

Scanning for Liquid samples

·         Place the liquid sample put on UATR crystal by using syringe.

·         Go to sample table and select standard as well as sample which put on the crystal.

·         Go to <measurement> Select <Scan> to Started Scanning.

·         Click on sample table on top right side and click to save.

Comparison of sample spectra with standard spectra or to overlay the spectra

·         Select the spectra to be compared with standard spectra by clicking on < file > and Select the spectra to be compared with standard spectra by clicking on < file > and    then on <open> and selecting the required spectra from opened window & click on open and let it be displayed on screen.

·         Go to <setup Menu> Select < compare> to open a setup window in bottom side to   select <setup compare reference> and Add Standard and Tick (√) on Correlation for result Pass/Fail.

·         Go to sample table to select respective Sample which compare with standard.

·         Go to <Process>Select <compare> to comparing Started.

Data Printing

·         Go to Sample View from the left side assistant bar to select the data for printing.

 

·         Go to <file menu> select < report> select < template> Select threw Browse< compare template> and press enter.

·         Feed your user name and password and press enter.

·         Take final print out by clicking on print.

·         Enter all necessary entry in Usage log book of instrument.

Calibration

Calibration Frequency: Quarterly

Calibration perform in three Parts:

1.            EP Resolution Performance Test

2.            EP wave number verification test

3.            Signal to Noise Ratio Test

Process

·         Place the normal base accessory carefully.

·         Go to <Setup menu> select <instrument Verification> a setup window is open at the bottom side.

·         Go to setup instrument Verification.

·         Go to instrument check, deselect the Abscissa, ordinate, noise and ASTM and select MIR Pharmacopoeia and then select European test.

·         Go to report and select both Show report automatically & Print report automatically.

·         Go to <Measurement menu> Select < instrument check> then click <instrument Verification>.

·         Instrument Verification-EP Resolution Performance Window is open.

·         Read the instruction carefully then click scan.

·         Follow the instruction onscreen carefully to complete the calibration.

·         SNR (Signal To Noise Ratio Test) (can be perform By Administrator only)

Process

Go to my computer to< C: drive> to <Program file> to <PerkinElmer> to < service IR> to <install instrument> and follow the instruction on screen run performance test.

Data printing

Go to my computer to< c: drive> to <Program file> to <PerkinElmer> to < service IR> to <Instrument Serial no.> to <Performance test> open this file and then print.

Action to be taken in case unsatisfactory results

·         If calibration of FTIR not match with acceptance criteria, inform Head/Designee-QC and affix the label ‘Out of Calibration’ to the instrument.

·         Do not use the Instrument if it is out of calibration.

·         Executive/Designee QC will inform to Maintenance Department.

·         Get the instrument corrected and recalibrate the instrument.

 

CALIBRATION OF FOURIER TRANSFORM INFRARED SPECTROPHOTOMETER (FTIR)

 

   Instrument ID No.                             :                                                            

Date of calibration                           :                                              

Next due date for calibration          :

Frequency of calibration                 : Quarterly

 

1. Wave number verification test:

Wave number

Observation

Acceptance Criteria

3060.0

 

3059.00 to 3061.00

2849.5

 

2848.50 to 2850.50

1942.9

 

1941.90 to 1943.90

1601.2

 

1600.20 to 1602.20

1583.0

 

1582.00 to 1584.00

1154.5

 

1153.50 to 1155.50

1028.3

 

1027.30 to 1029.30

2. Resolution:

    Absorbance mode:

1.)    Difference between the absorbance’s at 2849.5 cm-1 and 2870 cm-1  = …………...(> 0.33)

2.)    Difference between the absorbance’s at 1583 cm-1 and 1589 cm-1     = …………..... (> 0.08)

3. Signal to noise ratio

Observation: ………………………………. Limit (NLT 6500)

                                                                                 

Post a Comment

0Comments
Post a Comment (0)