C18 HPLC Column Washing Procedure

Rahul Kashyap
0

 

C18 HPLC Column Washing Procedure

C18 HPLC Column Washing Procedure
Photo Source 

Purpose

To describe the procedure for HPLC column washing.

Scope

This procedure is applicable for procedure for HPLC column washing

Procedure: C18 HPLC Column Washing Procedure

·         Reverse phase HPLC column washing procedure (Isocratic analysis method):

·         Flush the Column with 95 % water and 5 % Methanol for 15 minutes.

·         Flush the column with 5 % water and 95 % Methanol for 15 minutes.

·         Flush the column with 5 % water and 95 % Acetonitrile for 15 minutes.

·         Run the Gradient Program through HPLC column flow rate 1.0 ml/min for 45 minutes.

·         Washing method parameter are as described below;

Sr. No.

System parameter

Parameter

value

1.        

Data Acquisition

LC Stop time

45.01

2.        

Pump

Mode

Low pressure gradient

 

Flow

1.0000 ml/min

Line C Conc.

5 %

Line D Conc.

95 %

Pressure Max

449 kgf/cm2

3.        

Auto sampler

Auto sampler

Enable

 

Cooler temperature

25°C

4.        

Oven

Oven Temperature

40°C

5.        

Detector A

Lamp

Off

 

Gradient Programming is as mentioned below. C18 HPLC Column Washing Procedure

Time

Module

Command

value

0.01

Pumps

Solvent B Conc.

0

0.01

Pumps

Solvent C Conc.

5

0.01

Pumps

Solvent D Conc.

95

15.00

Pumps

Solvent B Conc.

0

15.00

Pumps

Solvent C Conc.

5

15.00

Pumps

Solvent D Conc.

95

15.01

Pumps

Solvent B Conc.

0

15.01

Pumps

Solvent C Conc.

95

15.01

Pumps

Solvent D Conc.

5

30.00

Pumps

Solvent B Conc.

0

30.00

Pumps

Solvent C Conc.

95

30.00

Pumps

Solvent D Conc.

5

30.01

Pumps

Solvent B Conc.

95

30.01

Pumps

Solvent C Conc.

0

30.01

Pumps

Solvent D Conc.

5

45.00

Pumps

Solvent B Conc.

95

45.00

Pumps

Solvent C Conc.

0

45.00

Pumps

Solvent D Conc.

5

45.01

Controller

Stop

 

               

·         Store the column with end fitting in respective column box.

·         Reverse phase HPLC column washing procedure (Gradient analysis method):

·         Select the solvent (Methanol/ Acetonitrile) as per major Constituents of mobile phase.

·         Flush the column with 5 % (Methanol/Acetonitrile) and 95 % water for 20 min.

·         Flush the column with 95 % (Methanol/Acetonitrile) and 5 % water for 20 min.

·         Run the Gradient Program through HPLC column flow rate 1.0 ml/min for 40 Minutes.

Washing method parameter are as described below. C18 HPLC Column Washing Procedure

Sr. No.

System parameter

Parameter

value

  1.  

Data Acquisition

LC Stop time

40.01

  1.  

Pump

Mode

Low pressure gradient

 

Flow

1.0000 ml/min

Line C Conc.

5 %

Line D Conc.

95 %

Pressure Max

449 kgf/cm2

  1.  

Auto sampler

Auto sampler

Enable

 

Cooler temperature

25°C

  1.  

Oven

Oven Temperature

40°C

  1.  

Detector A

Lamp

Off

Gradient Programming is given below.

Time

Module

Command

value

0.01

Pumps

Solvent C Conc.

5

0.01

Pumps

Solvent D Conc.

95

20.00

Pumps

Solvent C Conc.

5

20.00

Pumps

Solvent D Conc.

95

20.01

Pumps

Solvent C Conc.

95

20.01

Pumps

Solvent D Conc.

5

40.00

Pumps

Solvent C Conc.

95

40.00

Pumps

Solvent D Conc.

5

40.01

Controller

Stop

 

·        

Store the column with end fitting in respective column box.

·         Normal Phase HPLC Column Washing Procedure:

·         Flush the Column with HPLC grade IPA. C18 HPLC Column Washing Procedure

·         Run the HPLC grade IPA through HPLC column by 0.5ml/min to 1.5-ml/min flow where by pressure will not exceed above 3500 psi.

·         Flush the column with HPLC grade IPA for 45 minutes.

·         Washing method parameter are as described below.

                Sr. No.

System parameter

Parameter

value

  1.  

Data Acquisition

LC Stop time

45.01

  1.  

Pump

Mode

Low pressure gradient

 

Flow

1.0000 ml/min

Line D Conc.

100 %

Pressure Max

449 kgf/cm2

  1.  

Auto sampler

Auto sampler

Enable

 

Cooler temperature

25°C

  1.  

Oven

Oven Temperature

40°C

  1.  

Detector A

Lamp

Off

 

Gradient Programming is given below.

Time

Module

Command

value

0.01

Pumps

Flow

0.5000

15.00

Pumps

Flow

0.5000

15.01

Pumps

Flow

1.0000

30.00

Pumps

Flow

1.0000

30.01

Pumps

Flow

1.5000

45.00

Pumps

Flow

1.5000

45.01

Controller

Stop

 

 

Store the column with end fitting in respective column box.

Note: For Normal phase analysis Before Start the Batch file. purge D Line with the IPA for 10 minutes.