Calibration of Manometer | 15 steps in calibration

 

In this article you will learn the procedure of calibration of a manometer.The calibration of manometer can be carried out same as a pressure gauge calibration.

Calibration of Manometer 15 Steps in Calibration


·       Connect tube with low-pressure calibrator at ‘P1’ high-pressure side and connect other side to pressure generator.

Connect the unit under calibration with another nozzle of pressure generators. 

·       Leave the low-pressure side connection of pressure calibrator & unit under calibration to atmosphere.

·       Pressurize system by turning knob clockwise for pressure and anti-clockwise in case of vacuum.

·       Set various combination of pressure point from minimum to full span of manometer / low pressure gauges and digital indicator with transmitter and note down in record format 

·       Now gradually increase the pressure with the help of pressure generator by turning knob clockwise and take reading for upscale of low-pressure calibrator (master calibrator) & unit under calibration. 

·       Release pressure slowly by turning knob anticlockwise of pressure generator and take reading for down scale against low-pressure calibrator.

·       Note down reading for upscale and down scale in record format Annexure-I.

·       Calculate % variation of manometer/low pressure gauges and digital indicator with transmitter in upscale and down scale by using following formula.

·       Variation in % of span at upscale = [(X-Y1)/span] x 100 &

Variation in % of span at down scale = [(X-Y2)/span] x 100 &

·       Where X= low-pressure calibrator reading.

·       Where Y1 = upscale reading of manometer/low pressure gauges and digital indicator with transmitter under calibration.

·       Where Y2 = down scale reading of manometer/low pressure gauges and digital indicator with transmitter under calibration.

·       Note down variation in % of span in record of format Annexure-I.

·       Check the variation in % of span shall be within ± 3 %.

·       In case of out of calibration replace the manometer/low pressure gauges/digital indicator with transmitter.