Temperature
mapping report
PRE-REQUISTES
PROGRAM:
Temperature-monitoring
equipment, a monitoring device, a temperature data logger, or other such device that is suitable for its intended
purpose should be available and
used.
Document
details:
The
following documents should be verified and should be available before starting
the temperature mapping.
Standard
Operating Procedure of Temperature Mapping
Calibration
Certificates of the measuring Instruments (Data logger) and NABL certificate
Training
certificate of engineer of external agency
Training
record of Person involve in temperature mapping study
Calibration
details:
The
details of the calibrated instruments should be recorded.
Calibration
certificates should be attached with the report
VALIDATION/QUALIFICATION
TEST PROGRAMME:
Identification
of the area with rationale (Justification).
Storage
area identification and justification
Selection
of temperature mapping time and period:
The
study should be long enough to show any possible daily trend, so it should not
be less than 3 days (72 hours). Program the data loggers to acquire data every
10 minutes (Technical supplement to WHO TRS No. 961, 2011– Temperature mapping
of storage areas) in three different seasons as given below:
Sr.
No. Season Consider
Months
01 Winter
Nov- Feb. (with loaded condition)
02 Summer
Mar.-Jun.
(with loaded condition)
03 Rainy July-Oct. (with
loaded condition)
Establish
Data Loggers/Sensor Selection Criteria:
Data
Logger should be selected on the basis of following key features:
Data
Capacity:
Data Capacity determines how much data it can store. More data storage capacity
leads to store more data during its logging time.
Sample
Rate:
The frequency in which samples are taken. The loggers should be compatible to
provide data on user selectable sample rates.
Monitoring
Range and Accuracy:
Data Loggers shall have NIST-traceable 3-point
calibration certificate with a guaranteed error of not more than ±0.50C (for
temperature) at each calibration point, so that all anticipated temperature
extremes can be recorded.
Battery
Life:
Make sure the battery life is long enough to complete the mapping sessions.
Software: Make sure that the data
logger & software is capable to provide the intended data.
Identification
of number of Data Logger for Temperature Mapping with rationale (Justification):
Identify
the location and number of Data Logger
To
map the whole area.
Try
to use the Data Logger as many as possible to ensure a good distribution.
Make
sure, cover all storage areas and all storage levels, including top to bottom
or bottom to top, right to left or left to right, front to rear or rear to
front approach.
Considering
the Locations near doors, walls and ceilings may be more sensitive to exterior
conditions
Considering
the area include rack, door, window, supply and return grill, sun facing wall
with good distribution as well as obstruction of air.
Data
logger should be located at every 5 - 10 meters (WHO TRS 961 Annexure-9).
If
the ceiling height is 3.6 meters or less, position Data Logger directly above
one another at high and low level (e.g. one Data Logger at floor level, 1.2
meters and one Data Logger at 3.0 meters) Or as per required.
If
the ceiling height is greater than 3.6 meters, arrange Data Logger in vertical
arrays at the bottom, middle and top of the space. For instance, for storage
area six meters in height, three Data Loggers should be positioned in each grid
location heights of 1.8 meters, 3.6
meters and 5.4 meters.
Number
of required Data Logger for temperature mapping should be identified area wise.
Data
Logger Location with sketch diagram:
As
per the identified rationale Data Logger should be affixed according to the
prepared sketch diagram.
Performance
of the temperature mapping study:
Temperature
mapping study should be performed by following performance test:
Temperature
monitoring program for 72 Hours
Door
opening study
Power
failure study
Temperature
Monitoring study:
Introduction:
Temperature
mapping study is to evaluate the efficiency of area to maintain the temperature
within the acceptable limit and identify the Hot Point.
Procedure:
The controlling unit of the respective
area should be in running condition before one hour prior to execution of the
study and ensure that the controlling unit should be in running condition up to
the end of study. Controlling unit on time should be.
Set
temperature in controlling unit, so as to get temperature 25±2°C in the area.
Configure
the data-logger for recording frequency of every 10 minutes’ interval.
Place
the calibrated data logger within the room as per the define location with
proper label.
Start
the Temperature mapping with 10-minute interval for the total 72 hours.
Mark
the location in the area where data logger is to be placed (as per sketch
diagram) for mapping purpose and mention the location number.
Affix
the data logger at the locations identified for temperature mapping and review
the start time & end time from data logger report.
After
the programmed period, stop the Data Logger and remove the data logger from the
affixed location and download the data. Take the print out of downloaded data.
Attach
the downloaded printed data and graphical representation with report.
Collect
the graphical representation from the data logger placed in the mapped area for
verification of the uniformity of the temperature in the respective area and
same shall attached with the report.
Acceptance
criteria:
Limit:
Temperature should not exceed beyond the limit i.e. 25±2°C.
No
any excursion allowed beyond the limit i.e. temperature 25±2°C.
Identification
of hot spot & Cold Spot:
Identification
of hot & cold spot of the area and record the point.
Variation
of temperature at a single point Location:
Variation
of temperature at single point location should be identified.
Variation
of temperature at different Location point:
Variation
of temperature at different point location should be identified.
Length
of time of any temperature excursions:
Length
of time of any temperature excursion should be recorded.
Door
open Study:
Introduction: This study is to
evaluate the impact on temperature acceptance limit during opening of door in
routine working.
Procedure:
Ensure
that complete mapping has performed.
Open
the door and wait for 1, 5, 10 & 15 minutes and record the observation i.e.
door opening and closing time.
The
effect of the door opening on temperature shall be assessed and recorded after
every 10sec. for 1, 5, 10 & 15 minutes.
Acceptance
criteria:
Temperature at all the selected locations should be 25±2°C.
Power
Failure Study:
Introduction: Significance of power
failure study is to evaluate the Intervention of the power failure on
temperature of the room to maintain the temperature within the accepted limit
once if power cut exists.
Procedure:
Power
failure study to be carried out for 60 min.
Ensure
that the completion of mapping time i.e. 72 hours.
Power
failure challenge test should be performed by closing the controlling unit of
the area for 60 minutes.
Close
the controlling unit and record the closing time.
Record
the temperature at 10 min. interval for 60 min.
After
60 minutes start the controlling unit of the respective area and note down the time.
After
the programmed period, stop the Data Logger and remove the Data Logger from the affixed location and download /
extract the data. Take the print out of the data extract / downloaded from the
data-logger.
Acceptance
criteria:
Temperature at all the selected locations should 25±2°C.
REVALIDATION
CRITERIA:
Revalidation
shall be done in the following conditions
Change
in the design / layout of the area.
Any
change in the controlling unit, which can have an effect on the temperature of
the area.
Any
change in area configuration.