Brookfield viscometer calibration

Rahul Kashyap
2

Purpose:
The purpose of this SOP is to provide written guidelines for Calibration of Brook Field
Viscometer.
Procedure:
Frequency: Quarterly

  • Put the standard of 5 cps silicon fluid in a beaker, ensuring the level of spindle Up to groove.
  • Put the thermo-sensor in the sample.
  • Ensure the temp. of fluid is 250 C
  • Measure the viscosity at 100 RPM
  • Take two readings in cps to obtain average viscosity.
  • After finishing the work, the beaker should be thoroughly washed with distilled water.
  • Acceptance criteria: 5 cps + 1.0 % i .e.
  • Limit: 4.95 to 5.05 cps
  • Maintenance / Repairs:
  • When the instrument does not comply with the Acceptance criteria specified above, the instrument should be labeled “ out of calibration “ and should be repaired / serviced.

Precautions: 
  • Remove the Ultra Low adaptor (U. L.) & carefully clean the spindle using distilled water.  
  • Ensure that the spindle or any other part of the instrument is handled with 
  • extreme caution & the surfaces are not scratched as it can  disturb the calibration 
  • of the  instrument.
  • Maintain the cleanliness of instrument before & after use at each time.


Post a Comment

2Comments
  1. In 2003, this proposition was amended as Senate Bill (SB) 420 (Chapter 875, Statutes of 2003), which created dispensaries and a marijuana card program in LA. Moxie – Pear Cartridge

    ReplyDelete
  2. The first thing you need to do is find a reliable pharmacy in Canada. For every legitimate online pharmacy, there are 4 others which aren't. That means that as high as 80 percent of internet sites which offer Canada drug sales are registered in another country altogether! With odds like that, how will you ensure that you are receiving reliable and safe medications from pharmacies in Canada? hop over to these guys