Vacuum Leak Test | operation of vacuum leak tester
- Ensure instrument plug is properly connected to power socket.
- Display will show self-test process which mention “Connect Desiccator.”
- Check the connectivity of vacuum desiccator with instrument properly to avoid a leakage from tubes and its circumference lid. Display will also show “Press Enter (↵) Press ‘Enter’ and the instrument will perform self-test by creating vacuum.
- The instrument will check pump for its normal functioning. If pump function found ok the instrument shows ‘Pass’ status on the LCD screen.
- In case the function checks fail, place the desiccator lid again ensuring its firm and proper fitment on the desiccator. Restart the instrument.
- To operate the instrument press, enter (↵), the display will show “Test Run Menu” which consist “standard mode’ and “Admin mode” followed by date and time at the bottom.
- Place the sample can or bottle inside the desiccator wrapped in the tissue paper properly and cover sample top/ part/ area with tissue, where leakage needs to be checked.
- Enter the standard mode by pressing “Enter. The next display will show as following:
Vacuum Time
DEL Time (Delivery Time)
Hold Time
RET Time (Retention Time)
- Enter required values as per test, to be performed and press start to initiate the test.
-
- After completion of the test the instrument itself releases the vacuum according to the program.
-
- After testing wipe the desiccator for any leftover solvent with tissue paper.
Calibration
Take a calibrated vacuum gauge for calibration of instrument. Remove the tube from the instrument and connect to the vacuum gauge
Run the instrument as mentioned below in the table;
Set vacuum (mmHg) |
Hold time |
Tolerance (mmHg) |
150 |
30 sec |
135 - 165 |
300 |
30 sec |
270 - 330 |
500 |
30 sec |
450 - 550 |
- Frequency of calibration: Twelve months (for the timer, vacuum gauge and vacuum holding capacity).
Cleaning Procedure:
- Cleaning of leak test apparatus by using purified water and clean outer and inner surface with the help of the lint – free cloth.
-
- Place the apparatus for dryness for half an hour.
-
- Change the desiccator water twice in a week or as per required.