Humidity Calibration | Calibration of RH Controller
Calibration of relative humidity
hygrometers is generally performed in a humidity chamber.
The most used methods to generate
humidity
are the two pressure method, the split flow method and salt and water
solutions. The latter is one of the least expensive methods to perform a
calibration, this method
generally covering a wide range of relative humidity. According to, this method
we can obtain similar results to those obtained with a humidity chamber. Some
disadvantages with these methods are the possible contamination of the sensor,
the long times of
stabilization and the limitation
of only being able to calibrate the instrument to defined values.
This work describes the calibration
process of relative humidity hygrometers using as
moisture generation system, salt-water
solutions and validate the results with a system
of generation of humidity of two
pressures..
INTRODUCTION
The three most used methods to
generate humidity in the air are: the two pressures method
, the split flow method and salt-water
solutions. The mixture of salt and water allows to generate humidity relative
from about 3% to 98%. This wide range of humidity is useful for calibrate hygrometers
as well as for others Applications. The main advantage of this method is the
reduction in cost. The humidity value generated depends on the type of Salt,
table 1 shows the most used Salt solutions in the calibration.
Salt-water solutions and their
values Relative humidity reported in ASTM
Type of salt
|
HR(%)
|
t(ºC)
|
LiCI
|
11,3
|
25
|
MgCl2
|
32,8
|
25
|
NaCl
|
75,3
|
25
|
BaCl2
|
90
|
25
|
Due to the low cost of salt-water
solutions to generate moisture and based on the values to be considered as
fixed reference points, these are widely used by manufacturers, calibration
laboratories, and other users. However, recently, differences have been found
regarding the values reported by Greenspan , OIML and ASTM, so it is convenient to characterize
these solutions with a suitable technique or instrument (direct measurement of vapour
pressure, meter dew point temperature).
A disadvantage is that there is
still no reliable and repeatable calibration procedure based on salt solutions.
Among the most important aspects to consider in the preparation of salt
solutions are the following:
1) Material of the container
containing the aqueous solution,
2) The purity of the components
of the solution,
3) The preparation technique,
4) The volume and design of the
calibration chamber,
5) the measurement and control of
temperature inside the chamber 6) Characterization
The container should be made of a
non-hygroscopic material, resistant to corrosion. Teflon, stainless steel and
glass are commonly used.
The salt used must be analytical
reagent grade. The water used in the solution must be distilled or de-ionized.
A portion of salt must be
completely dissolved in water and maintain an undissolved amount (excess salt)
which can be 30% to 90% compared to the weighed sample.
Another important factor to
consider is the volume and design of the test chamber. The stabilization times,
the volume of water-salt to be used and the temperature and humidity gradients
depend on this.
The temperature and the variation
in the solubility, constitutes the main factor of influence in the solutions.
Temperature instability at ± 0.1 ºC, can cause relative humidity instabilities
of ± 0.5% RH of the generated value.
As already mentioned, depending
on the design and volume of the test chamber, it is possible that relative
humidity and temperature gradients are generated, therefore it is necessary to
carry out a characterization to determine these gradients and their effect on
uncertainty. The amounts of water and salt required were calculated from the
solubility of each of the salts at a temperature of 23 ºC.
Calibration:
Measuring
In this method, the measurand is
the resulting correction value of the instrument under calibration (IUC)
CR = [(HRp+CR´ )- (HR) IUC] + [δ (HR)t +
δ (HR)hist + δ
(HR)stb ]
Where: CR is the correction value
resulting from the IUC reading (in% relative humidity); (HRp + CR´) is the
corrected reading of the pattern; CR 'is the correction value of the standard
reading; (HR) IUC is the instrument reading under calibration; (HR) t is the
temperature effect correction value; (HR) camera is the stability correction
value in the camera; (HR) hysteresis is the correction value for hysteresis
effect
Process
The calibration procedure is based on the “Technical Guide
on traceability and measurement uncertainty in the calibration of relative
humidity hygrometers” and is summarized in the following diagram. The selected
calibration points were described in table
Figure shows a diagram summarizing the calibration
procedure.
The salts were characterized before starting the calibration
process. The volume of the container was considered to be small, so the effect
of gradients is neglected and only stability in relative humidity is
considered.
UNCERTAINTY CALCULATION
The sources of uncertainty considered in this calibration
system are
Pattern uncertainty Up
Camera uncertainty U camera
Uncertainty of the readings U reading
Temperature Uncertainty U Temperature
Hysteresis uncertainty U Hysteresis
The standard uncertainty is that reported on your
calibration certificate. Chamber uncertainty includes stability and gradients
in temperature and humidity.
The uncertainty of the readings is given by:
Where HR1 is the value of relative humidity measured upwards;
HR2 is the value of relative humidity measured downwards.
This method is adequate if the necessary care is taken into
account such as:
The purity of the materials.
The preparation of the solutions.
Characterization of the calibration zone.
Temperature control.
Humidity measurement with calibrated instrument.