HPLC Calibration Parameters

Rahul Kashyap
0


HPLC Calibration Parameters
Image Source:Wikipedia

HPLC Calibration Parameters
Following Parameters are used during calibration of HPLC:
·         Flow rate accuracy test
·         Retention time reproducibility:
·         Gradient composition:
·         Calibration of injector:
·         Injection precision:
·         Injector & detector linearity:
·         Carry over test
·         Injection accuracy:
·         Wave length accuracy:
·         Temperature accuracy:
·         Sampler cooler temperature accuracy
·         Calibration of pda detector (spd – m20a):
·         Lamp energy
CALIBRATION of HPLC System
(SHIMADZU LC2030C- PROMINACE -i)/(PDA-Detector)
(1)    FLOW RATE ACCURACY TEST:
Mobile Phase: water                                                                   Stop Watch ID. No: …………………….
        Stop Watch Calibration Due on: ……………………….
Sr. No
Set flow rate
Time in minutes
Observed vol. at 250C =
{(Wt. of V. flask +water) –(Wt. of V. flask)}/0.997
Observed flow in ml =
  Observed Vol.
  Observed time
 Limit
( ±1%)
Remarks.

1
0.5 ml/min
20 minute
Observed Time =



± 0.005
Complies/
Does            not comply
2
1.0 ml/min
10 minute
Observed Time =



± 0.01
Complies/
Does
not comply
3
2.0 ml/min
5 minute
Observed Time =



± 0.02
Complies/
Does
not comply
4
2.5 ml/min
4 minute
Observed Time =



± 0.025
Complies/
Does
not comply

                                                                                                                                      
                                                                                                                      Remarks:  Pass/ Fail

        (2) RETENTION TIME REPRODUCIBILITY:
Caffeine
Batch No.                                                     Make:                                 Used before date:                                                         
Column ID:
Chromatographic conditions:
HPLC Condition                    : Column C18 150 x 4.6 mm, 5µ.
Detector                                  : 272 nm (for caffeine)
Injection volume                     : 20 µl.
Mobile phase               : Methanol: water (70:30)
Flow rate                                 : 1.0 ml/min.
Take ………………..mg of Caffeine and dilute to 100 ml with mobile phase. Further dilute 2 ml to 50 ml with mobile phase.
Injection No.
Retention time
Limit
1




% RSD not more than 1.0
2

3

4

5

6

Mean

SD

% RSD



                             



                                                                                                        Remarks: Pass /Fail

    




      (3) GRADIENT COMPOSITION:

Composition
Composition Using A & C
 Composition Using B & D

Limit
m.v
% of composition
m.v
% of composition
C100






C90





89.0 to 91.0
C50





49.0 to 51.0
C10




9.0 to 11.0



     Calculation: for channel A & C
     Actual concentration of 10 % (C) = ---------------- x100

     Actual concentration of 50 % (C) = ---------------- x100

     Actual concentration of 90 % (C) = ---------------- x100

    Calculation: for channel B & D
   
    Actual concentration of 10 % (D) = ---------------- x100

    Actual concentration of 50 % (D) = ---------------- x100

       Actual concentration of 90 % (D) = ---------------- x100
                                                                                                                                   


                                                                                                              Remarks: Pass /Fail


       (4) Calibration of Injector:
      (A).  INJECTION PRECISION:                                                                                 
Injection No.
Area
Limit
1




% RSD not more than 2.0
2

3

4

5

6

Mean

SD

% RSD



                                                                                                                          Remarks:  Pass/ Fail
      
       (B) INJECTOR & DETECTOR LINEARITY:

Injection No.
Injection Volume
                      Area
Limit








1
       5

2
     10

3
     20

4
     30

5
     40

6
     50


7
     100

Correlation  coefficient

NLT 0.999

                                                                                                                                       
                                                                                                                          Remarks:  Pass/ Fail








      (C) CARRY OVER TEST: (Use 250 mcg/ml caffeine standard solution and mobile phase as a
                                                     blank)
Sr. No.
Mean area of solution
 Area of caffeine in blank injection
% Carry Over
Limit






NMT 0.05%
                                                                                                                                           

                                                                                                            Remarks: Pass /Fail

      (D) INJECTION ACCURACY:
Sr. No
Weight of vial + water
( W1  in mg )
Set volume
µl
Number of injections
‘n’
Wt. of vial + water in mg (W2)
( after 6 injections )
Avg. amount of injected water in µl
Acceptance
value
in µl


20
6


19.6 to 20.4 µl

                                                              
                                                                                                                     Remarks: Pass /Fail

           Calculation:
                                                                             W1- W2
          Average Volume of Injection = ------------------                      
                                                                           0.997 X n

     (5) WAVE LENGTH ACCURACY:
    Column ID: _______________________
   (i) From wave length Maxima (202 nm – 210 nm):
S. No.
Wavelength
Absorbance
Observed maxima at
Limit
1.         
202


205 ± 2 nm
2.         
203

3.         
204

4.         
205

5.         
206

6.         
207

7.         
208

8.         
209

9.         
210


                                                                                                                     Remarks: Pass /Fail

         (ii) From wave length Minima (240 nm – 248 nm):

S. No.
Wavelength
Absorbance
Observed Minima at
Limit
1.         
240


245 ± 2 nm
2.         
241

3.         
242

4.         
243

5.         
244

6.         
245

7.         
246

8.         
247

9.         
248


          
                                                                                                                            
                                                                                                                            Remarks: Pass /Fail
 


    (iii) From wave length (268 nm – 276 nm):

S. No.
Wavelength
Absorbance
Observed maxima at
Limit
1.         
268


273 ± 2 nm
2.         
269

3.         
270

4.         
271

5.         
272

6.         
273

7.         
274

8.         
275

9.         
276

                                        
                                                                                                                                                     Remarks: Pass /Fail                                                                                                                                                                                           

   (6) Temperature Accuracy:
       Oven Temperature:
        Thermometer ID: _________________              Calibration certificate No.: ____________

        Calibration Valid up to: _____________

Sr. No
 Set  Temperature  
Displayed Temp.
Observed temp of thermometer
Limits:
1
25 ° C


24 to 26 ° C
2
40 ° C


39 to 41 ° C
3
60 o C


59 to 61 o C

                                                                                                                                                     
                                                                                                                          Remarks: Pass /Fail





     (7) Sampler cooler temperature accuracy:
       Thermometer ID No. …………………………….                          Calibration due date ……………..……….
Sr. No
 Set  Temperature  
Displayed Temp.
Observed temp of thermometer
Limits
1
4o C


   3°C to 5°C
2
10° C


   9°C  to 11°C
3
40 ° C


  39°C to 41°C

                                                                                                                       Remarks: Pass /Fail
  (8) CALIBRATION OF PDA DETECTOR (SPD – M20A):
  (A) Detector performance:
       PDA Detector Precision:
   Preparation of Caffeine stock solution (200 mcg):
  Weigh accurately ..........  mg Caffeine in 100 ml volumetric flask; add 60 ml mobile phase to dissolve and make up the volume with same.
    Preparation of Caffeine solution (20 mcg):
     Transfer .............. ml of Caffeine stock solution in 50 ml volumetric flask and make up and mix with          mobile phase.
        Solution Used:
Injection
No.
Retention
Time
Area
Response
Observed
 % RSD
Acceptable
% RSD
Test Result
1.



NMT 1.0 % For
Retention  Time



Pass/fail
2.


3.


4.



NMT 2.0 % For
Area
5.


6.


Avg.



SD



 RSD



                                                                                                                         Remarks: Pass /Fail
                                                                                                                                        
       (B) PDA Detector Linearity:

Preparation of Caffeine stock solution (200 mcg):

Weigh accurately ................ mg of Caffeine in 100 volumetric flasks.

Preparation of Caffeine stock solution (10 mcg):

Transfer .............. ml of Caffeine Stock solution in 100 ml volumetric flask.

Preparation of Caffeine stock solution (20 mcg):

Transfer .............. ml of Caffeine Stock solution in 50 ml volumetric flask.

Preparation of Caffeine stock solution (40 mcg):

Transfer .............. ml of Caffeine Stock solution in 10 ml volumetric flask.

Preparation of Caffeine stock solution (80 mcg):

Transfer .............. ml of Caffeine Stock solution in 25 ml volumetric flask.

Preparation of Caffeine stock solution (100 mcg):

Transfer .............. ml of Caffeine Stock solution in 10 ml volumetric flask.

S. No.
Concentration of Caffeine
Area Response
Observed
      ( r )
Acceptable
     ( r )
Test
 Results
1.
10 mcg






NLT 0.999




Pass / Fail
2.
20 mcg

3.
40 mcg

4.
80 mcg

5.
100 mcg

                                                               
                                                                                                                     Remarks: Pass /Fail


    (9) LAMP ENERGY:

Observation

Limit: NLT 400

                                                                                                                 Remarks: Pass /Fail

        Conclusion: Instrument meets / does not meet the acceptance criteria.


Post a Comment

0Comments
Post a Comment (0)