OBJECTIVE
To
ensure that the desired temperature and pressure required to sterilize the
load
is achieved at different places in the load placed in an autoclave.
SCOPE
This
protocol is intended to prove with a high degree of confidence that the
sterilization process is consistent, reproducible, under control and able to
supply a final product which fulfils all the sterilisation specification
attributes. This will be demonstrated by monitoring the critical sterilisation
parameters, according to what established in this protocol.
A
concurrent validation approach has been chosen to perform three consecutive runs
for each consisting empty load, full load and third full load with biological
indicators according to the validation protocol. All the equipment, procedures
and personnel involved in the validation process will be the same as those used
for normal production batches. Procedures and documentation will comply with the
current GMP and Global Quality Standards.
The
validation of the autoclave shall be divided into three parts as follows:
Part A :
Thermal Studies
Part B :
Biological Studies
Part C :
Correlation of Thermal and Biological studies to determine the
Fo delivered by
the process.
Part A - Thermal Studies (Heat
penetration & heat distribution)
The
studies shall be performed to validate the "Performance" of the
sterilizer for delivering the desired
lethality at all the points in the sterilizer equally and
uniformly.
The
studies, therefore shall include the following aspects: -
Use
of Thermocouples (T type) min. 12 nos.
for thermal mapping of the
steam
sterilizer using a data logger. It will include
- Empty Run
- Maximum Load
Since
these are trial runs for 3220D type granule with small batch size Minimum Load
validation runs are not performed. However, these will be performed on regular
batch size.
- Monitoring the pressure inside the autoclave using a calibrated sensor mounted on the sterilizer.
- Check for correlation between 'RTD’ s already in place and additional thermocouples together with the gauges mounted for correct operating parameters
- Studying the effect of various loads and loading patterns on the thermal distribution and heat penetration.
Determination
of the following points for critical evaluation: -
a.
'Slow to heat' points
b. 'Cold' point
c. 'Most-difficult-to-sterilize’ point.
(Note:
All relevant calibration documents will be made available.)
Part
B - Biological Studies
This
aspect involves the challenging of the sterilizer with an appropriate biological
load to evaluate the efficacy of the sterilization process.
Under
the present condition where low temperature sterilization is being carried out
viz. at 112.5 ºC for 50 mins. the most appropriate organism would be “Bacillus subtilis"
ATCC: 35021 or ‘5230' with a D - value of 0.3 mins. Based on the data obtained
from the thermal studies the Biological Indicators (BI) would be placed
appropriately to evaluate the lethality delivered by the process. After the
exposure of the BI’s it will take 48 hrs to obtain the results. (As per diagram
attached)
(Certificate
of Analysis will be made available mentioning D - value and the empirically
derived survival/kill time and Fo values)
Part
C - Correlation of Thermal and Biological Studies to determine the
Process
Fo
This
will be the final step wherein the Process Fo value would be derived from the
thermal data.
Note:
If sufficient lethality is not being achieved at all the points in the
sterilizer uniformly, relevant recommendations will be given to achieve
the same.
Procedure
of Validation: - For each autoclave following runs shall be taken.
- Empty Run - to setup the thermocouples at various locations and run the cycle.
- Maximum Load - with the thermocouples kept inside the bottles (by cutting the bottle at the top and sealing the bottle with the help of cellophane tape) at various locations and run the cycle. In the third cycle of full load, BI’s shall be placed adjacent to thermocouple (inside the bottles) to verify lethality at the locations.
(generally
the maximum thermal load will be employed as a "WORST
CASE
MODEL.
In
this validation since batch size is 5000 L for 1000 ml, 1200 L for 500 ml and
600 L for 100 ml, number of bottles produced may not be sufficient for maximum
l load, * so for making full load bottles from another batch shall be taken.
This is being done especially for this validation studies. In normal validation
studies or production cycles such condition shall not prevail.
Instruments:
Instruments:
Digital
Temperature Indicator & controller:
Calibrated
Pressure
Gauge: Calibrated
Calibrated
Data logger
Calibrated
RTD’S – Min. Twelve numbers.
PROCESS
DESCRIPTION
BRIEF
DESCRIPTION OF OPERATION
- Calibrated temperature sensors shall be connected to autoclave and then to data logger before starting autoclave validation.
- Data logger shall be started and temperature shall be recorded at interval of 1 minute during each sterilization cycle.
- Pressure and temperature on the dial shall be noted manually. (indicated temperature and pressure)
- Biological validation shall be done on third full load of sterilization of each autoclave. Ampoules of biological indicators shall be kept inside the bottles by cutting the bottle at the top and then by sealing it with the cellophane tape in which the thermocouples are kept in previous cycles. One ampoule of biological Indicator shall be kept outside the autoclave as a positive control during each sterilization cycle.
- For negative control one ampoule of negative control shall be incubated along with these ampoules.
- After completion of sterilization cycle, ampoules of spores shall be collected and incubated at 35-39 °C for 72 hours. And results shall be recorded in the separate protocol.
ACCEPTANCE
CRITERIA:
- The holding time as determined from the measured temperature shall not be less than 50 minutes for Minimum Load and Maximum Load.
- During the holding time -
- The measured temperature shall be within the average sterilization temperature of 112.5 to 114.5 °C.
- The measured temperature in each load item does not fluctuate by more than 2 °C. The indicated chamber pressure shall be within 0.05 bar of set pressure.
- The
F0 value for each load should be greater than or equal to 8 and the Sterility
Assurance Level (SAL) should be 10` 6. for Minimum Load and Maximum Load.
- Ampoules of spores exposed to sterilization shall not show any evidence of growth in all the sterilization cycles during specified incubation period i.e. colour of the ampoule shall remain as it is. (red colour)
- Positive control should show growth in terms of turbidity and change in colour from red to yellow within 24 hours of incubation.
- Negative control should not show any evidence of growth i.e. red colour of the ampoule shall remain red only.
- If the growth resulted from organism other than the test organism, the test is inconclusive and should be repeated.
AT
THE END OF THE CYCLE:
The
temperature sensors shall have remained in position.
The
items containing sensors should be intact.
Printouts
of data logger shall be checked for appropriate temperature.
EVALUATION
AND CERTIFICATION:
Data
obtained from data logger and Sterility reports of biological indicators shall be
evaluated by Quality Assurance Department.
After
confirming the satisfactory result, Quality Assurance head shall certify the
validation.
CORRECTIVE
ACTION IN CASE OF VALIDATION FAILURE:
The
Quality Assurance Head shall decide on corrective action to be initiated.
The
corrective action shall be based on investigation of various factors affecting
the study like:
Points
related to Autoclave,
Steam
leakage, Valve adjustment, Air pockets, Condition of heating element
Calibration
of instruments (in consultation with Head of Maintenance department.)
An
investigation report shall be prepared jointly by Quality Assurance and
Maintenance specifying the corrective actions to be taken. The validation
process shall be repeated after taking the corrective actions. The autoclave shall not be used till the autoclave is
revalidated.
Frequency
of Steriliser Validation: - Once in a
Year.