Operation and Calibration of Moisture Balance

Rahul Kashyap
0

Purpose
This SOP describes the procedure for operation and calibration of Moisture Balance.
Scope
This procedure is applicable for operation and calibration of Moisture Balance located at ABC Limited
Procedure :
Operation
  • Clean the instrument with clean and dry cloth.
  • Ensure that the instrument is calibrated.
  • Turn ON the main switch followed by the ‘ON/OFF’ switch,.
  • Switch on the moisture analyzer.
  • Select drying program & set the drying parameters.
  • Open the sample chamber and place an unused sample pan on the pan support.
  • Tare the sample pan and select tare and confirm.
  • Distribute approx. 2 g sample evenly on the sample pan.
  • Close the sample chamber.
  • Current moisture loss and elapsed time are displayed on the display.
  • Calibration

  • Daily Verification of Weighing Pan
  • Turn on the analyzer.
  • Select CAL button for calibration.
  • Confirm CAL & Press Enter.
  • Confirm again when Pb is shown. Press Enter.
  • Tare the weighing System & Press Enter.
  • Select CAL again.
  • Confirm CAL & Press Enter.
  • The prompt for calibration weight is displayed on the display
  • Now open the hinged cover.
  • Place the 30 g weight on the weighing system. Minus sign indicates weight value to low, Plus Sign indicates weight valve to high & No sign indicates weight value ok.
  • The weight unit symbol (g) is displayed at the end of adjustment.
  • Unload the analyzer. Close the cover.
  • Quit Calibration.
  • Acceptance Criteria: 30 ± 0.02 %
  • Monthly Calibration of Moisture Content
  • Clean the instrument with clean and dry cloth.
  • Turn ON the main switch followed by the ‘ON/OFF’ switch,.
  • Switch on the moisture analyzer.
  • Select drying program & set the drying parameters.
  • Now set the temperature to 1500C.
  • Open the sample chamber and place an unused sample pan on the pan support.
  • Tare the sample pan and select tare and confirm.
  • Distribute approx 2-3 g of Disodium Tartarate evenly on the sample pan.
  • Close the sample chamber.
  • Current moisture loss and elapsed time are displayed on the display.
  • Acceptance Criteria: 14-17 % at 1500 C

Precaution
  • Handle the instrument carefully and properly.
  • Do not handle the aluminium panel with oily or greasy fingers.
  • Do not use abrasive or corrosive substances to clean the aluminium panel.